Q: When there is an alert, does the logger send an update to the cloud immediately and send the email alert to the customer or does it wait till the hourly upload?

...

A: See: Set Temperature Warnings

 

Q: How many recipents can receive alarms?

...

If the RF signal is less then 50% for a logger, then the logger might be sending data irregularly.

 

Q. In some cases I don’t get Data Alarm (Data Warning – Lost connection to a logger) sent even though no measurement data is logged in the database and no measurements can be seen in the Controlant cloud multigraph or reports. Why?

Q. In some cases I can see temperature measurements from a logger and the logger is connected to the gateway but no data can be seen in the multigraph or report. Why?

Answer

Both of these questions can have the same root cause and is not related to any defect in the system. The reason is related current behaviour of the system or to so called 01 and 03 data package.

What is 01 and 03 Data packages?

01 data package               A data packed that is sent from a logger and includes connection information to the gateway and temperature data. 01 data temperature information is NOT stored in the data base and NOT used to generate graph or report

03 data package               A data package that is sent from a logger when it has been verified with 01 data package that the logger is in good connection with a gateway. The 03 data package includes measured temperature data that is stored in the data base.

Problem occurs in bad connection

When a logger is in a bad connection to a gateway (less than 50% signal strength) the gateway can receive the 01 data packages and show the measured temperature in the User interface. But the connection is not good enough for the logger to receive confirmation from the gateway so the logger doesn’t send the 03 data package to the gateway. There is a “signal strength” icon in the logger information bar that indicates the signal strength

In cases like this, the user can see, in the User Interface, that the Logger in connection to the gateway (1), receives temperature measurement (2) and shows signal strength (3) even though the signal strength is very low.