NoteFor continuity of measurements, a logger replacing another logger in a re-calibration or warranty process should be associated with the the same measuring point. In the re-calibration process, the logger that is being replaced and the logger that is replacing the old logger should be placed together in the same conditions for a couple of hours. It is not possible to view the temperature readings of the new logger before it's added to a measuring point |
...
3. Select MODIFY from the drop-down menu
...
Info | |
---|---|
title | NOTEThe ID of the logger currently associated with the measuring point will be highlighted in grey on the list of sensors: |
...
4. Find the ID of the new logger on the list of the sensors and click on it, it will be highlighted in blue
5. Click on SAVE
...
Info |
---|
NOTEFrom the moment when the new ID was associated with the measuring point, the temperature data will be logged from a new logger |
...